Development of a Monte Carlo Simulation Model for Varian ProBeam Compact Single-Room Proton Therapy System using GEANT4

File
Publisher
Florida Atlantic University
Date Issued
2020
EDTF Date Created
2020
Description
Proton therapy with pencil beam scanning technique is a novel technique to treat cancer patients due to its unique biophysical properties. However, a small error in dose calculation may lead towards undesired greater uncertainties in planed doses. This project aims to create a simulation model of Varian ProBeam Compact using the GEANT4 Monte Carlo simulation tool kit.
Experimental data from the first clinical ProBeam Compact system at South Florida Proton Therapy Institute was used to validate the simulation model. A comparison was made between the experimental and simulated Integrated Depth-Dose curves using a 2%/2mm gamma index test with 100% of points passing. The beam spot standard deviation sizes (s!) were compared using percent deviation. All simulated s! matched the experimental s! within 2.5%, except 70 and 80 MeV. The model can be used to develop a more comprehensive model as an independent dose verification tool and further investigate dose distribution.
Note

Includes bibliography.

Language
Type
Extent
57 p.
Identifier
FA00013547
Rights

Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.

Additional Information
Includes bibliography.
Thesis (M.S.)--Florida Atlantic University, 2020.
FAU Electronic Theses and Dissertations Collection
Date Backup
2020
Date Created Backup
2020
Date Text
2020
Date Created (EDTF)
2020
Date Issued (EDTF)
2020
Extension


FAU

IID
FA00013547
Organizations
Person Preferred Name

String, Shawn

author

Graduate College
Physical Description

online resource
57 p.
Title Plain
Development of a Monte Carlo Simulation Model for Varian ProBeam Compact Single-Room Proton Therapy System using GEANT4
Use and Reproduction
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
http://rightsstatements.org/vocab/InC/1.0/
Origin Information

2020
2020
Florida Atlantic University

Boca Raton, Fla.

Physical Location
Florida Atlantic University Libraries
Place

Boca Raton, Fla.
Sub Location
Digital Library
Title
Development of a Monte Carlo Simulation Model for Varian ProBeam Compact Single-Room Proton Therapy System using GEANT4
Other Title Info

Development of a Monte Carlo Simulation Model for Varian ProBeam Compact Single-Room Proton Therapy System using GEANT4