Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance

File
Publisher
Florida Atlantic University
Date Issued
2016
EDTF Date Created
2016
Description
The purpose of this research is to accurately match the calculation environment, i.e. the treatment planning system (TPS) with the measurement environment (using a 2-D diode array) for lung Stereotactic Body Radiation Therapy (SBRT) patient-specific quality assurance (QA). Furthermore, a new phantom was studied in which the 2-D array and heterogeneities were incorporated into the patient-specific QA process for lung SBRT.
Dual source dual energy computerized tomography (DSCT) and single energy computerized tomography (SECT) were used to model phantoms incorporating a 2-D diode array into the TPS. A water-equivalent and a heterogeneous phantom (simulating the thoracic region of a patient) were studied. Monte Carlo and pencil beam dose distributions were compared to the measured distributions. Composite and individual fields were analyzed for normally incident and planned gantry angle deliveries. The distributions were compared using γ-analysis for criteria 3% 3mm, 2% 2mm, and 1% 1mm.
The Monte Carlo calculations for the DSCT modeled phantoms (incorporating the array) showed an increase in the passing percentage magnitude for 46.4 % of the fields at 3% 3mm, 85.7% at 2% 2mm, and 92.9% at 1% 1mm. The Monte Carlo calculations gave no agreement for the same γ-analysis criteria using the SECT.
Pencil beam calculations resulted in lower passing percentages when the diode array was incorporated in the TPS. The DSCT modeled phantoms (incorporating the array) exhibited decrease in the passing percentage magnitude for 85.7% of the fields at 3% 3mm, 82.1% at 2% 2mm, and 71.4% at 1% 1mm. In SECT modeled phantoms (incorporating the array), a decrease in passing percentage magnitude were found for 92.9% of the fields at 3% 3mm, 89.3% at 2% 2mm, and 82.1% at 1% 1mm.
In conclusion, this study demonstrates that including the diode array in the TPS results in increased passing percentages when using a DSCT system with a Monte Carlo algorithm for patient-specific lung SBRT QA. Furthermore, as recommended by task groups (e.g. TG 65, TG 101, TG 244) of the American Association of Physicists in Medicine (AAPM), pencil beam algorithms should be avoided in the presence of heterogeneous materials, including a diode array.
Note

Includes bibliography.

Language
Type
Extent
171 p.
Identifier
FA00004744
Additional Information
Includes bibliography.
Dissertation (Ph.D.)--Florida Atlantic University, 2016.
FAU Electronic Theses and Dissertations Collection
Date Backup
2016
Date Created Backup
2016
Date Text
2016
Date Created (EDTF)
2016
Date Issued (EDTF)
2016
Extension


FAU

IID
FA00004744
Organizations
Person Preferred Name

Curley, Casey Michael

author

Graduate College
Physical Description

application/pdf
171 p.
Title Plain
Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
Use and Reproduction
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
http://rightsstatements.org/vocab/InC/1.0/
Origin Information

2016
2016
Florida Atlantic University

Boca Raton, Fla.

Physical Location
Florida Atlantic University Libraries
Place

Boca Raton, Fla.
Sub Location
Digital Library
Title
Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
Other Title Info

Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance