Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery

File
Publisher
Florida Atlantic University
Date Issued
2022
EDTF Date Created
2022
Description
Dosimetric uncertainty in very small (< 2 x 2 cm2) photon fields is notably higher that has created research questions when using small-field virtual cone with variable multileaf collimator (MLC) fields. We evaluate the efficacy of the virtual cone with a fixed MLC field for stereotactic radiosurgery (SRS) of small targets such as trigeminal neuralgia.
We employed a virtual cone technique with a fixed field geometry, called fixed virtual cone (fVC), for small target radiosurgery using the EDGE (Varian Medical Systems, Palo Alto, CA) linac. The fVC is characterized by 0.5 cm x 0.5 cm high-definition MLC field of 10 MV flattening filter-free (FFF) beam defined at 100 cm SAD, while jaws are positioned at 1.5 cm x 1.5 cm. A spherical dose distribution equivalent to 5 mm cone was generated by using 10–14 non-coplanar partial arcs. The dosimetric accuracy of this technique was validated using the SRS MapCHECK (Sun Nuclear Corporation, FL) and the EBT3 (Ashland Inc., NJ) film based on absolute dose measurements. For the quality assurance (QA), 10 treatment plans for trigeminal neuralgia consisting of various arc fields at different collimator angles were analyzed retrospectively using 6 MV and 10 MV FFF beams, including the field-by-field study (n = 130 fields). Dose outputs were compared between the SRS MapCHECK measurements and Eclipse treatment planning system (TPS) with Acuros XB algorithm (version 16.1). In addition, important clinical parameters of 15 cases treated for trigeminal neuralgia were evaluated for the clinical performance. Moreover, dosimetric (field output factors, dose/MU) uncertainties considering a minute (± 0.5–1.0 mm) leaf shift in the field defining fVC, were examined from the TPS, SRS diode (PTW 60018) measurements, and Monte Carlo (MC) simulations.
Note

Includes bibliography.

Language
Type
Extent
128 p.
Identifier
FA00013958
Rights

Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.

Additional Information
Includes bibliography.
Dissertation (Ph.D.)--Florida Atlantic University, 2022.
FAU Electronic Theses and Dissertations Collection
Date Backup
2022
Date Created Backup
2022
Date Text
2022
Date Created (EDTF)
2022
Date Issued (EDTF)
2022
Extension


FAU

IID
FA00013958
Organizations
Person Preferred Name

Neupane, Taindra

author

Graduate College
Physical Description

application/pdf
128 p.
Title Plain
Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery
Use and Reproduction
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
http://rightsstatements.org/vocab/InC/1.0/
Origin Information

2022
2022
Florida Atlantic University

Boca Raton, Fla.

Place

Boca Raton, Fla.
Title
Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery
Other Title Info

Efficacy of the virtual cone method using fixed small multi-leaf collimator field for stereotactic radiosurgery