Comparison of treatment plans calculated using ray tracing and Monte Carlo algorithms for lung cancer patients having undergone radiotherapy with cyberknife

File
Publisher
Florida Atlantic University
Date Issued
2014
EDTF Date Created
2014
Description

The purpose of this research is to determine the feasibility of introducing the Monte Carlo (MC) dose calculation algorithm into the clinical practice. Unlike the Ray Tracing (RT) algorithm, the MC algorithm is not affected by the tissue inhomogeneities, which are significant inside the chest cavity. A retrospective study was completed for 102 plans calculated using both the RT and MC algorithms. The D95 of the PTV was 26% lower for the MC calculation. The first parameter of conformality, as defined as the ratio of the Prescription Isodose Volume to the PTV Volume was on average 1.27 for RT and 0.67 for MC. The results confirm that the RT algorithm significantly overestimates the dosages delivered confirming previous analyses. Correlations indicate that these overestimates are largest for small PTV and/or when the ratio of the volume of lung tissue to the PTV approaches 1.
Note

Includes bibliography.

Language
Type
Extent
83 p.
Identifier
FA00004316
Additional Information
Includes bibliography.
Thesis (M.S.)--Florida Atlantic University, 2014.
FAU Electronic Theses and Dissertations Collection
Date Backup
2014
Date Created Backup
2014
Date Text
2014
Date Created (EDTF)
2014
Date Issued (EDTF)
2014
Extension


FAU

IID
FA00004316
Organizations
Person Preferred Name

Pennington, Andreea

author

Graduate College
Physical Description

application/pdf
83 p.
Title Plain
Comparison of treatment plans calculated using ray tracing and Monte Carlo algorithms for lung cancer patients having undergone radiotherapy with cyberknife
Use and Reproduction
Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
http://rightsstatements.org/vocab/InC/1.0/
Origin Information

2014
2014
Florida Atlantic University

Boca Raton, Fla.

Physical Location
Florida Atlantic University Libraries
Place

Boca Raton, Fla.
Sub Location
Digital Library
Title
Comparison of treatment plans calculated using ray tracing and Monte Carlo algorithms for lung cancer patients having undergone radiotherapy with cyberknife
Other Title Info

Comparison of treatment plans calculated using ray tracing and Monte Carlo algorithms for lung cancer patients having undergone radiotherapy with cyberknife