Dosimetric Consequences From Minimal Displacements In Abpi With Savi Applicators

File
Publisher
Florida Atlantic University
Date Issued
2015
EDTF Date Created
2015
Description
Purpose: To highlight the importance of provide proper solid immobilization initially and in every
treatment in ABPI with brachytherapy. Materials and Methods: 100 patients receiving brachytherapy
treatments in ABPI using the Savi applicators were considered for the study. The CT scans used in the
initial plan was compared with every scan obtained before each treatment. Each of these scans were exported in the planning system and registered with the initial plan. Dosimetric evaluations were
performed with their consequences to the ribs and the skin surface. Results: Making the dosimetric
comparison for the critical points on the ribs and skin due to very small changes in the interfractionation
position revealed dramatic differences in the maximum dose to these critical organs. The
cavity's volume manifested changes between fractions as well as the distances to the two critical
organs. Therefore the maximum dose manifested variance between 10 and 32 in both of the sites
Conclusions: This study demonstrates that using CT scan before each treatment will minimize the risk
of delivering undesired high doses to the critical organs. This reveals the urgent need of increasing and
improving the immobilization methods when treating ABPI with Savi. In 30 of the cases re-planning was
necessary between fractions therefore we conclude that in each case the treatment and planning teams
must be prepared for re-plan as needed.
Note

The Sixth Annual Graduate Research Day was organized by Florida Atlantic University’s Graduate Student Association. Graduate students from FAU Colleges present abstracts of original research and posters in a competition for monetary prizes, awards, and recognition.

Language
Type
Genre
Extent
1 p.
Identifier
FA00005872
Additional Information
The Sixth Annual Graduate Research Day was organized by Florida Atlantic University’s Graduate Student Association. Graduate students from FAU Colleges present abstracts of original research and posters in a competition for monetary prizes, awards, and recognition.
FAU Student Research Digital Collection
Date Backup
2015
Date Created Backup
2015
Date Text
2015
Date Created (EDTF)
2015
Date Issued (EDTF)
2015
Extension


FAU

IID
FA00005872
Organizations
Attributed name: Graduate College
Person Preferred Name

Chandrasekara, Shereen
Physical Description

application/pdf
1 p.
Title Plain
Dosimetric Consequences From Minimal Displacements In Abpi With Savi Applicators
Use and Reproduction
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Origin Information

2015
2015
Florida Atlantic University

Boca Raton, Fla.

Physical Location
Florida Atlantic University Libraries
Place

Boca Raton, Fla.
Sub Location
Digital Library
Title
Dosimetric Consequences From Minimal Displacements In Abpi With Savi Applicators
Other Title Info

Dosimetric Consequences From Minimal Displacements In Abpi With Savi Applicators