MEK (2-butanone) and THF (tetrahydrofuran) were found in high concentrations in our laboratory's water outlets 6 mo after PVC (polyvinyl chloride) pipe installation. Subsequent analysis confirmed that the solvents from the PVC pipe cement used to join the tubing were leaching into our water supply. Water samples were taken at various residence times in the pipe to observe the solvents' leaching kinetics.

Language
Type
Genre
Extent
5 p.
Identifier
3176699
Additional Information
MEK (2-butanone) and THF (tetrahydrofuran) were found in high concentrations in our laboratory's water outlets 6 mo after PVC (polyvinyl chloride) pipe installation. Subsequent analysis confirmed that the solvents from the PVC pipe cement used to join the tubing were leaching into our water supply. Water samples were taken at various residence times in the pipe to observe the solvents' leaching kinetics.
This manuscript is an author version with the final publication available at http://www.springerlink.com and may be cited as: Wang, T. C., & Bricker, J. L. (1979). 2-Butanone and tetrahydrofuran contamination in the water supply. Bulletin of Environmental Contamination and Toxicology, 23(1), 620-623. doi:10.1007/BF01770014
Florida Atlantic University. Harbor Branch Oceanographic Institute contribution #141.
Date Backup
1979
Date Text
1979
DOI
10.1007/BF01770014
Date Issued (EDTF)
1979
Extension


FAU
FAU
admin_unit="FAU01", ingest_id="ing10691", creator="creator:BCHANG", creation_date="2011-09-20 07:27:12", modified_by="super:FAUDIG", modification_date="2014-02-11 14:04:19"

IID
FADT3176699
Issuance
single unit
Organizations
Person Preferred Name

Wang, Tsen C.

creator

Physical Description

pdf
5 p.
Title Plain
2-Butanone and tetrahydrofuran contamination in the water supply
Origin Information

Springer-Verlag.
1979
single unit
Title
2-Butanone and tetrahydrofuran contamination in the water supply
Other Title Info

2-Butanone and tetrahydrofuran contamination in the water supply