Publisher
Florida Atlantic University
Description
Sickle Cell Disease (SCD) is a genetic disease that affects approximately 100,000 people in the USA and millions worldwide. The disease is defined by a mutation in hemoglobin, the red blood cell’s oxygen carrying component. Under hypoxic (low oxygen) conditions, the mutated hemoglobin (known as HbS) polymerizes into rigid fibers that stretch the cell into a sickle shape. These rigid cells can occlude blood vessels and cause an individual immense pain. Currently, no point-of-care devices exist in the market for assisting those with SCD. Using microfluidics with custom designed portable impedance measuring hardware we can achieve label-free in vitro analyses of SCD rheology.
This dissertation presents two impedance-based devices for finger-prick volume blood testing, including a microflow cytometer for SCD diagnostics and a vaso-occlusion tester for monitoring blood flow activities. First, the microflow cytometer is validated by measuring the electrical impedance of individual cells flowing through a narrow microfluidic channel. Cellular impedance is interpreted by changes in subcellular components due to oxygen association-dissociation of hemoglobin, using an equivalent circuit model and Multiphysics simulation. Impedance values of sickle cells exhibit remarkable deviations from normal blood cells. Such deviation is quantified by a conformity score, which allows for measurement of SCD heterogeneity, and potentially disease severity. Findings from this study demonstrate the potential for SCD screening via electrical impedance. Second, a vaso-occlusion tester is validated by measuring the impedance response of blood flow within a microfluidic mimic of capillary bed.
Rights
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Person Preferred Name
Dieujuste, Darryl
author
Graduate College
Title Plain
MICROFLUIDIC ELECTRICAL IMPEDANCE TECHNOLOGY FOR POINT-OFCARE ASSESSMENT OF SICKLE CELL DISEASE
Use and Reproduction
Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Title
MICROFLUIDIC ELECTRICAL IMPEDANCE TECHNOLOGY FOR POINT-OFCARE ASSESSMENT OF SICKLE CELL DISEASE
Other Title Info
MICROFLUIDIC ELECTRICAL IMPEDANCE TECHNOLOGY FOR POINT-OFCARE ASSESSMENT OF SICKLE CELL DISEASE