Measurements of output factors for small photon fields up to 10 cm x 10 cm

File
Publisher
Florida Atlantic University
Date Issued
2013
EDTF Date Created
2013
Description
Field output factors (OF) for photon beams from a 6 MV medical accelerator
were measured using five different detectors in a scanning water phantom. The
measurements were taken for square field sizes of integral widths ranging from 1 cm to
10 cm for two reference source-to-surface distances (SSD) and depths in water. For the
diode detectors, square field widths as small as 2.5 mm were also studied. The photon
beams were collimated by using either the jaws or the multileaf collimators. Measured
OFs are found to depend upon the field size, SSD, depth and also upon the type of beam
collimation, size and type of detector used. For field sizes larger than 3 cm x 3 cm, the
OF measurements agree to within 1% or less. The largest variation in OF occurs for jawsshaped field of size 1 cm x 1cm, where a difference of more than 18% is observed.
Note

Includes bibliography.

Language
Type
Extent
78 p.
Identifier
FA0004003
Additional Information
Includes bibliography.
Thesis (M.S.)--Florida Atlantic University, 2013.
Date Backup
2013
Date Created Backup
2013
Date Text
2013
Date Created (EDTF)
2013
Date Issued (EDTF)
2013
Extension


FAU

IID
FA0004003
Issuance
single unit
Organizations
Person Preferred Name

Bacala, Angelina

author

Graduate College
Physical Description

Online Resource
78 p.
Title Plain
Measurements of output factors for small photon fields up to 10 cm x 10 cm
Use and Reproduction
http://rightsstatements.org/vocab/InC/1.0/
Origin Information

2013
2013
Florida Atlantic University
single unit
Physical Location
Florida Atlantic University Digital Library
Sub Location
Boca Raton, Fla.
Title
Measurements of output factors for small photon fields up to 10 cm x 10 cm
Other Title Info

Measurements of output factors for small photon fields up to 10 cm x 10 cm