Model
Digital Document
Publisher
Florida Atlantic University
Description
In this work, we have developed a robust daily quality assurance (QA) system for pencil-beam scanning (PBS) dosimetry. A novel phantom and multi-PTV PBS plan were used in conjunction with the Sun Nuclear Daily QA3 multichamber detector array to verify output, range, and spot position. The sensitivity to detect change in these parameters with our designed tests was determined empirically. Associated tolerance levels were established based on these sensitivities and guidelines published in recent American Association of Physics in Medicine (AAPM) task group reports. The output has remained within the 3% tolerance and the range was within ±1mm. Spot position has remained within ±2mm. This daily QA procedure is quick and efficient with the time required for setup and delivery at less than 10 minutes.